简介
实验室主要功能是固结以及非固结岩芯超声波速度、孔隙度、电阻率及渗透率等岩石物理参数的测试、分析工作。
主要仪器
岩石物理分析室始建于2015年,目前拥有一套SCMS-F高/低温高压岩芯多参数测量系统及数套配套设备。
SCMS-F高/低温高压岩芯多参数测量系统
本仪器由同济大学海洋地质国家重点实验室和西南石油大学国家重点实验室联合研制,属新型全自动岩芯测量系统,拥有测量参数多,精度高等特点。可用于模拟天然气水合物在高压低温环境下的生成过程。同时也可在高温高压条件下,针对固结及非固结岩芯的超声波速度、孔隙度、渗透率等岩石物理参数进行测量,其主要技术指标为:
固结岩芯部分:
实验温度:室温~150℃误差:±1℃
围压压力:1~90MPa误差:±1.5%
孔隙压力:0~10 MPa误差:±0.25%
岩芯直径:25mm 岩芯长度:3~7cm
气体孔隙度测量范围:0.1~30PU误差:±0.25PU
气体渗透率测量范围:~2000(µm2)误差:±15%
液体渗透率测量范围:~1000(µm2)误差:±5%
电阻率测量范围:0.1~1000Ω·m误差:±5%
纵波测量范围:150~1000µs/m误差:±2%
横波测量范围:150~1000µs/m误差:±5%
非固结部分:
实验温度:-6℃~常温误差:±1℃
轴向压力:1~40MPa误差:±1.5%
孔隙压力:0~40 MPa误差:±0.25%
岩芯直径:6 (cm) 岩芯长度:0~2cm
气体渗透率测量范围(低内压):~1000(µm2)误差:±15%
液体渗透率测量范围:~1000(µm2)误差:±5%
电阻率测量范围:1~1000Ω·m误差:±5%
纵波测量范围:>=800m/s误差:±2%
横波测量范围:>=450m/s误差:±5%
The Rock Physics Lab
Introduction
The rock physics lab has been built since 2015 and by now there is a SCMS-F multi-parameter measure system with several auxiliary apparatuses. In this lab, the main works are focus on the measurement and analyses of ultra-elastic velocities, porosity, resistivity and permeability in well-consolidate and unconsolidated rock samples.
The SCMS-F system is manufactured by the state key lab of marine geology in Tongji university and oil and gas reservoir geology and exploitation in Southwest petroleum university. It is designed to measure multi rock physical parameter in high accuracy with an auto-control system. It can be used to simulating gas hydrate in high pressure and low temperature. Furthermore, the rock physical parameters, such as ultra-elastic velocities, porosity, permeability and resistivity, can also be measured for well-consolidate and unconsolidated core samples in high pressure and temperature. The following parts are the key parameters used in core sample tests.
For well-consolidate core samples:
Temperature:room temperature ~150℃ relative error :±1℃
Confining pressure:1~90MPa relative error:±1.5%
Pore pressure:0~10 MPa relative error:±0.25%
Core diameter:25mmCore length:3~7cm
The range for porosity measurement:0.1~30PU relative error:±0.25PU
The range for gas-induced permeability measurement:~2000(µm2)
relative error:±15%
The range for fluid-induced permeability measurement:~1000(µm2)
relative error:±5%
The range for resistivity measurement:0.1~1000Ω·m
relative error:±5%
The range for longitudinal wave :150~1000µs/m relative error:±2%
The range for shear wave:150~1000µs/m relative error:±5%
For unconsolidated samples:
Temperature:-6℃~room temperaturerelative error:±1℃
Axial Pressure:1~40MParelative error:±1.5%
Pore Pressure:0~40 MParelative error:±0.25%
Core diameter:6cmCore length:0~2cm
The range for gas-induced permeability measurement:~1000(µm2)
relative error:±15%
The range for fluid-induced permeability measurement:~1000(µm2)
relative error:±5%
The range for resistivity measurement:1~1000Ω·mrelative error:±5%
The range for longitudinal wave:>=800m/s relative error:±2%
The range for shear wave:>=450m/s relative error:±5%